Positron emission tomography (PET) and single photon emission computed tomography (SPECT), the two most common nuclear medicine medical imaging technologies, are based on the detection of gamma radiation. In the case of PET, radiotracers are β+ emitters, which upon an annihilation with an electron produce pairs of gamma rays. In the case of SPECT radiotracers, gamma photons are directly emitted upon a nuclear de-excitation of one of the short-lived excited nuclei.
SPECT and PET detectors are both based on the same principle technology: a scintillator coupled through a light guide to a photo-detector. The principal aim of the scintillator is to stop gamma rays and convert the absorbed energy into visible scintillation light, which are subsequently collected by the photo-detector and converted into an electrical pulse.
There are several types of photo-detectors, and photomultiplier tubes (PMT) and photodiodes (PD) i.e. silicon photomultipliers (SiPM) or avalanche photodiodes (APD) are the most popular. PMTs consist of a vacuum tube with a photocathode at the entrance, a series of dynodes at successively high voltages, and an anode at the end of the PMT. The photocathode detects the optical photons emitted by the scintillator, and, by photoelectric effect, electrons are emitted in the direction of the dynodes, through the focusing electrodes, and accelerated by the existing electric field. When electrons interact with the dynodes, these are refocused to the next dynode, and new electrons are generated from the dynode. The amplification of electrons is exponential and results in an exceptional gain, when electrons reach the anode. PMTs have excellent timing and gain properties. However, these devices are bulky, delicate, operate at high voltages, and the electrons-focusing stage along the dynodes does not function in the presence of electromagnetic fields.
PDsSiPMs are considered the semiconductor analogue of PMTs. A photon travelling through silicon may interact with a bound electron and transfer its energy. This energy transfer causes the
electron to move to the valence shell and generate an electron-hole pair. PDsSiPMs are based on a densely packed array of avalanche photodiodes (p-n junction) operating in Geiger mode, i.e., in reverse-bias voltage over the breakdown voltage. All photodiodes are connected through a polysilicon quenching resistor and are read in parallel. When a photon from the scintillator interacts with a photodiode, it causes a breakdown discharge that results in a photoelectron pulse of high gain. The output of the PDSiPM is the sum of all the pulses from every photodiode that detected a photon. PDsSiPMs have gain comparable to PMTs operating at low voltages, are very compact and fast, and can function in the presence of electromagnetic fields [1].
From the physical principle’s perspective, the basic differences between PET and SPECT cameras, besides the different radiotracers to which they are sensitive, originate in the dual gamma-emission per nuclear decay for PET, while radionuclides used in SPECT emit only one gamma-photon per nuclear decay. This difference determines the different geometries used in PET and SPECT systems. PET systems consist of a ring of detectors, while SPECT systems may consist of one to three rotating detectors, but normally there are only two detectors fixed at 90° or 180° on a rotating gantry.
The detection principle in PET relies on electronic coincidence detection also known as electronic collimation) with high detection efficiency, as compared to SPECT, which requires a high density (usually made of lead or tungsten) physical collimator to determine the direction of the incoming photons. Collimators are designed to absorb gamma-photons coming from non-perpendicular directions, which implicitly limits the detection efficiency. Collimators can have several structures, like parallel-hole, multi-pinhole, multi-slit or fan-beam structure, among others. Parallel-hole collimators consist of an array of parallel holes which can be large or small depending on whether high sensitivity or spatial resolution is aimed. Pinhole collimators consist of an array of cone-shaped holes and have a better trade-off between sensitivity and resolution than parallel-hole collimators, but these usually cover a small field of view. While pinhole collimators magnify the object that is being imaged, their sensitivity is dependent on the distance to the object.
Another difference between PET and SPECT cameras resides in the different scintillators used for each modality. The scintillators are especially designed to provide the best properties in a range of energies depending on radionuclide. Gamma-photons produced from PET radiotracers have an energy of 511 keV, while the energy from gamma-photons emitted from SPECT radiotracers varies depending on the radionuclide and range from 110 keV to >350 keV. The energy of the gamma photons, and how they interact with the SPECT collimators, has a significant impact on the spatial resolution. Some common radionuclides for PET and SPECT are shown in Tables 1 and 2, respectively.
|
Radionuclide |
Positron energy (Mean/Max) |
Positron range (Mean/Max) |
Half life |
|
11C 18F |
0.39/0.96 MeV 0.28/0.65 MeV |
1.2/4.2 mm 0.7/2.5 mm |
20.4 min 110.0 min |
|
13N 68Ga |
0.49/1.20 MeV 0.84/1.90 MeV |
1.8/5.5 mm 2.9/9.0 mm |
10.0 min 67.7 min |
|
15O 11C |
0.73/1.73 MeV 0.39/0.96 MeV |
3.0/8.4 mm 1.2/4.2 mm |
2.0 min 20.4 min |
|
18F 15O |
0.28/0.65 MeV 0.73/1.73 MeV |
0.7/2.5 mm 3.0/8.4 mm |
110.0 min 2.0 min |
|
82Rb |
1,.5/3.3 MeV |
16.0/7.5 mm |
1,25 min |
|
68Ga 11C 13N |
0.84/1.90 MeV 0.39/0.96 MeV 0.49/1.20 MeV |
2.9/9.0 mm 1.2/4.2 mm 1.8/5.5 mm |
67.7 min 20.4 min 10.0 min |
|
Radionuclide |
Photon energy |
Half life |
|
99mTc |
141 keV (89%) |
6.02 hours |
|
|
|
|
|
111In
|
171 keV (91%) and 245 keV (94%) |
2.80 days |
|
123I |
159 keV (83%) |
13.22 hours |
|
131I |
364 keV (81%), 637 keV (7 %) |
8,03 days |
|
177Lu |
208 keV (10.4%) and 113 keV (6.8%) |
6.70 days |
|
|
|
|
|
Radionuclide |
Energy |
Half life |
|
99mTc |
141 keV (89%) |
6.02 hours |
|
177Lu |
208 keV (10.4%) and 113 keV (6.8%) |
6.70 days |
|
123I |
159 keV (83%) |
13.22 hours |
|
111In |
171 keV (91%) and 245 keV (94%) |
2.80 days |
|
155Tb |
87 keV (32%) and 105 keV (25%) |
5.32 days |
Scintillators in SPECT are generally continuous slabs decoded by an underlying array of photo-detectors. Upon an interaction, light propagates in the scintillator, and the intrinsic spatial resolution degrades with the scintillator thickness. To overcome the dependence between spatial resolution and scintillator thickness, there are already some systems with pixelated scintillators [2]. In the case of PET scintillators, they generally consist of an array of crystals separated by reflector material to confine the light in each crystal when an interaction takes place. The crystal size is directly related to the reconstructed image spatial resolution, but small crystals may require a significant number of photo-detectors to accurately decode the gamma interaction position, and this increases the total cost of the PET system.
Scintillators have several physical properties that affect the efficiency, time, and energy resolutions of PET/SPECT detectors and cameras. The density and atomic number determine the efficiency of the detector, which impacts the detector sensitivity. The light yield determines the number of optical photons generated from each gamma-interaction within some range which affects the time and energy resolution. The decay time determines how fast the luminescence signal decays after an excitation. Finally, the emission wave-length is the wave-length of the produced optical photons, which needs to be compatible with the wave-length to which the photo-detector is sensitive to. A fraction of the produced optical photons is converted into electrons in the photo-detector depending on the quantum efficiency of the photo-detector. The scintillator properties are directly related to intrinsic detector properties, namely the sensitivity, energy resolution, and time resolution. Other aspects to consider from scintillators are the hygroscopicity, how difficult is to cut them and how finely they can be cut, and radiation hardness among others. Some typical scintillators used in PET and SPECT are shown in table 3.
As mentioned before, PET systems consist of a ring of detectors to detect the pairs of gamma photons from a positron-electron annihilation. In the case of SPECT cameras, usually one to three heads located in a rotating gantry are used to detect the gamma photons. PET and SPECT systems are usually attached to a computer tomography (CT) system to acquire anatomical information without moving the patient and accurately overlay the functional information from the PET or SPECT on a perfectly aligned anatomical reference [3].
|
Scintillator |
Technology |
Density |
Light yield |
Emission |
Decay time |
|
LSO(Ce) |
PET |
7.40 |
26,000 |
410 |
40 |
|
BGO |
PET |
7.13 |
8,200 |
480 |
300 |
|
La3Br(Ce) |
PET |
5.29 |
61,000 |
358 |
35 |
|
NaI(Tl) |
SPECT/PET |
3.57 |
38,000 |
415 |
230 |
|
CsI(Tl) |
SPECT |
4.51 |
54,000 |
550 |
1,000 |
Typical SPECT systems consist of two gamma cameras mounted on a rotating gantry. However, there are several SPECT systems with singular configurations especially designed for specific organs. Most attention has been directed towards designing dedicated systems for cardiac SPECT to study myocardial perfusion.
There are several dedicated cameras based on cadmium-zinc-telluride (CZT) technology [5]. CZT detectors directly convert gamma radiation to electrical signals, thereby circumventing the need of a scintillator and improving the energy and spatial localization accuracy of gamma interactions. There are conventional whole-body SPECT systems based on CZT such as the Discovery NM/CT 670 (GE Healthcare, Waukesha, WI, USA), but there are also several dedicated cardiac SPECT systems based on CZT technology with a variety of collimation systems: stationary multi-pinhole, enhancing spatial resolution (Discovery 5(3)70c, GE Healthcare, Waukesha, WI, USA) [6], and rotational parallel-hole (D-SPECT, Spectrum Dynamics, Haifa, Israel) which enhance sensitivity [7].
In the GE Discovery system, each hole is focused on one pixel from the CZT detector (2.46x2.46 mm), thereby eliminating the light sharing problem. The system has 19 cameras focused on the heart, fully covers the heart and its surrounding tissue, and produces a high resolution image in a volume of 19 cm diameter, which needs to be centred in the heart (guided by an automatic algorithm) to achieve optimum performance.
The Spectrum Dynamics D-SPECT system consists of 9 rotating cameras placed in a curved configuration. The collimator is thinner (21.7 mm) and has larger holes (2.26 mm) than conventional collimators used in whole-body SPECT, and this results in a significantly larger solid angle and, hence, sensitivity. As in the case of the stationary multi-pinhole system, each hole is focused in one single CZT pixel, which is 2.50x2.50 mm.
There are also approaches to develop dedicated brain SPECT systems focused on Parkinson’s and similar neurodegenerative diseases, which, rather than a two-head structure, have a ring geometry similar to that from PET systems, like the Inspira HD system (Neurologica, Boston, USA). The main aim of these dedicated brain systems is to achieve a spatial resolution in the range of ~3 mm compared to 7-12 mm obtained with whole-body systems [8]. The INSERT research project also requires special mention, whose aim is to implement a high-resolution MR-compatible brain SPECT insert based on mini-slit-slat collimators coupled to a monolithic CsI scintillator on a SiPM array [9].
Conventional PET is based on detectors consisting of a scintillator coupled to a photo-detector based on analogue electronics such as PMTs and PDsSiPMs. Optical photons produced in a scintillator following an interaction by a gamma-photon are emitted isotropically through the scintillator. A fraction of these optical photons eventually reaches the entrance of the photo-detector through an optical guide (grease, glue, or similar). Each photon interacting in a micro-cell generates an avalanche of carriers resulting in a pulse of current containing 105-106 electrons and lasting a few nanoseconds. For accurate determination of energy and time information of the detected photons, the output signal is usually amplified through a low noise pre-amplifier, and the pulse shape is enhanced and cleaned before the analogue signal is digitized.
A current trend in the development of PET detectors is to perform the digitization process at an early point in the electronic chain, like CZT detectors that perform the digitization upon gamma photon detection. However, CZT detectors are currently used only in SPECT [10] and small animal PET systems [11].
Another trend is performing the digitization directly in the photo-detector upon detection of optical photons. The digital photon counter (DPC), developed by Philips (Philips Healthcare, Best, Netherlands), was first presented in 2009 [12]. With digital PDSiPMs, the pulse generated in each SiPM diode is immediately digitized upon generation eliminating any signal variations due to electronic noise and temperature fluctuations. Digital PDsSiPMs have several benefits over analogue SiPMs, which are briefly discussed below.
In the case of analogue PDsSiPMs, the time stamp of an interaction is obtained from a discriminating leading edge as the sum of all the photodiodes. In the case of digital SiPMs, the signal from each photodiode is digitized, effectively functioning as a photon counter, and potentially providing the time stamp of each detected photon from the scintillator, which has been shown to improve the system coincidence time resolution (CTR) [13].
Several efforts have investigated the time performance of digital SiPMs coupled to scintillators of different sizes reaching a CTR of ~130 ps [14].
In order to identify the position following a gamma interaction with pixelated scintillators based on light sharing, centre of gravity is commonly used. In addition to the great performance in timing resolution, using the digitized SiPM outputs, advanced algorithms can be used to improve the interaction position accuracy [15].
The digitization process close to the SiPM results in outstanding sensitivity and volumetric spatial resolution. Commercial PET/CT Systems today are equipped with digital SiPM detectors. Only a few still use the traditional PMT-Based analog detectors. Standard PET/CT systems employ detector ring assemblies reaching up to about 30cm Axial Field of View. These are referred to as SAFOV-PET Systems. Since about a decade, there are PET/CT systems with configurable detector ring assemblies reaching up to about 200cm Long Axial Field of View, referred to as LAFOV PET/CT systems.
There have been significant advances in the development of novel PET detectors in the last few years. Depth of interaction by using multi-layer or monolithic scintillators to increase the spatial resolution uniformity across the field of view has been the subject of several studies including whole-body scanners[16], but especially in dedicated brain[17] and small animal PET scanners[18] due to the close proximity of the PET detectors and the field of view.
The change from PMTs to PDsSiPM detectors has opened the possibility of increasing the amount of light collected in the photo-detector and the solid angle coverage to produce a significant boost in detected radiation and hence sensitivity and count rate. This increase in sensitivity enables a number of improvements in PET imaging, namely reduced injected radiotracer (facilitating the scan of children and pregnant women), short scan times, and improved image quality and, if used in LAFOV-PET Systems, a further push in sensitivity coming with genuinely new options in studying pharmacokinetic and -dynamic processes throughout the body simultaneously as well as very late imaging time point post application of respective radioligands with regard to the half life of the radionuclide used, respectively.
Overall, most technical advances in clinical systems have been focused on developing fast detectors to obtain high CTR, mainly driven by time-of-flight PET applications. Increased CTR has been shown to produce significantly high image quality and increasing the image contrast, improving small lesions detectability, and increasing the spatial resolution.
CTR relies on the scintillator, photo-detector and electronic chain, and these three parts have been the focus of different research studies. All PET manufacturers have put special effort on improving the CTR of their scanners and setting the minimum at 250 ps for the moment in a clinical PET/CT scanner. There is an active community working on new scintillating materials, new doping ions, and new photo-detectors, and some research groups are working on improving the timing and efficiency properties of scintillators and photo-detectors with the aim of 10 ps on the horizon[19].
Apart from imaging equipment, a number of non imaging radiation detecting, measuring and quantifying equipment is used in nuclear medicine.
First and foremost this is activity meters, often referred to as “dose calibrators”. This is a crucial instrument for the safe and accurate application of the desired amounts of activity of diagnostic and therapeutic radiopharmaceuticals to patients. In this context it is important to name it so the denotation instructs the correct application, that is measuring, quantifying, dispensing a defined/prescribed amount of radioactivity, rather than (whatever) dose. As we in Nuclear Medicine actually use the term ‘dose’ in another context, that is the description of absorbed radiation doses by the distribution of certain amounts (or fractions thereof) of activity in volumes/tissues/organs it is very important to make this distinction in language use and denotation.
Activity meters typically comprise a cylindrically shaped shielded ionization chamber operating in proportional mode surrounding as much as possible of the activity aliquot to be measured. By calibration of countrates from particular windows of energy of the radiation to actual activity, a quantification of radioactivity (in terms of Bq) is possible. This calibration is done for various geometries of aliquots (i.e. syringes, vials etc.) and different entities of emitted radiation (i.e. alpha-, beta- and gamma radiation) and different energies of those.
Moreover, well counters are available and are basically activity meters too but designed to be much more sensitive to very low amounts of radioactivity to be measured. Technically, they are constructed similar to activity meters but have much more sensitive radiation detectors that are designed to give a linear response in lower orders of magnitude than activity meters do. Activity meters typically give a linear response on the order of several one digit MBq to some two digit GBq whilst well counters do so in orders of magnitude of several Bq to some three to four digit kBq. Well counters are used to quantify (very low) radioactivity or concentration of radioactivity in small samples of excretions (i.e. urine, saliva or faeces) or whole blood and (after centrifugation) in plasma, respectively.
Finally, there are radiation detecting probes that are used to localize and only relatively quantify radioactivity by showing an impulse or count rate (i.e. ips or cps) to give an impression of the local distribution and uptake of radioactivity. Usually scintillation detectors are in use and are designed to detect a variety of radiations and energies of mostly gamma and beta radiation. Handheld designs, nowadays wireless coupled to the audiovisualization device, are most commonly in use and are designed to be disinfectable and/or sterilizable so they can be used in-situ to detect tissue that has taken up radioactivity/tracer. The most widely spread use case is sentinel lymph node detection accompanying planar and SPECT imaging by actually localizing the beforehand imaged nodes and, thereby, supporting surgical interventions. Probes designed to be fitted into endoscopic/robotic surgical equipment are also available.
1. Roncali E, Cherry SR. Application of silicon photomultipliers to positron emission tomography. Ann Biomed Eng 2011;39:1358–77. https://doi.org/10.1007/s10439-011-0266-9.
2. Sabet H, Bläckberg L, Uzun-Ozsahin D, El-Fakhri G. Novel laser-processed CsI:Tl detector for SPECT. Med Phys 2016;43:2630–8. https://doi.org/10.1118/1.4947294.
3. Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. Journal of Nuclear Medicine : Official Publication, Society of Nuclear Medicine 2000;41:1369–79. PMIDs: 10945530.
4. Handbook of Nuclear Medicine and Molecular Imaging for Physicists Instrumentation and Imaging Procedures, Volume I; Physics in Nuclear Medicine Book • Fourth Edition • 2012
5. Kincl V, Drozdová A, Vašina J, Panovský R, Kamínek M. Kadmium-zinek-telluridové SPECT kamery - nové perspektivy nukleární kardiologie. Cor et Vasa 2015;57:e214–8. https://doi.org/10.1016/j.crvasa.2015.01.001.
6. Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, Volokh L. A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential. Eur J Nucl Med Mol Imaging 2010;37:1887–902. https://doi.org/10.1007/s00259-010-1488-z.
7. Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, et al. A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med 2009;50:635–43. https://doi.org/10.2967/jnumed.108.060020.
8. Stam MK, Verwer EE, Booij J, Adriaanse SM, de Bruin CM, de Wit TC. Performance evaluation of a novel brain-dedicated SPECT system. EJNMMI Phys 2018;5:4. https://doi.org/10.1186/s40658-018-0203-1.
9. Busca P, Occhipinti M, Trigilio P, Cozzi G, Fiorini C, Piemonte C, et al. Experimental Evaluation of a SiPM-Based Scintillation Detector for MR-Compatible SPECT Systems. IEEE Transactions on Nuclear Science 2015;62:2122–8. https://doi.org/10.1109/TNS.2015.2481184.
10. Keidar Z, Raysberg I, Lugassi R, Frenkel A, Israel O. Novel Cadmium Zinc Telluride Based detector General Purpose Gamma Camera: Initial Evaluation and Comparison with a Standard Camera. J Nucl Med 2016;57:259–259.
11. Abbaszadeh S, Gu Y, Reynolds PD, Levin CS. Characterization of a sub-assembly of 3D position sensitive cadmium zinc telluride detectors and electronics from a sub-millimeter resolution PET system. Phys Med Biol 2016;61:6733–53. https://doi.org/10.1088/0031-9155/61/18/6733.
12. Frach T, Prescher G, Degenhardt C, de Gruyter R, Schmitz A, Ballizany R. The digital silicon photomultiplier — Principle of operation and intrinsic detector performance. 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC), 2009, p. 1959–65. https://doi.org/10.1109/NSSMIC.2009.5402143.
13. Fishburn MW, Charbon E. System Tradeoffs in Gamma-Ray Detection Utilizing SPAD Arrays and Scintillators. IEEE Transactions on Nuclear Science 2010;57:2549–57. https://doi.org/10.1109/TNS.2010.2064788.
14. Liu Z, Pizzichemi M, Auffray E, Lecoq P, Paganoni M. Performance study of Philips digital silicon photomultiplier coupled to scintillating crystals. J Inst 2016;11:P01017–P01017. https://doi.org/10.1088/1748-0221/11/01/P01017.
15. Vandenberghe S, Mikhaylova E, D’Hoe E, Mollet P, Karp JS. Recent developments in time-of-flight PET. EJNMMI Physics 2016;3:3. https://doi.org/10.1186/s40658-016-0138-
16. Berg E, Roncali E, Kapusta M, Du J, Cherry SR. A combined time-of-flight and depth-of-interaction detector for total-body positron emission tomography. Med Phys 2016;43:939–50. https://doi.org/10.1118/1.4940355.
17. Camarlinghi N, Belcari N, Cerello P, Sportelli G, Pennazio F, Zaccario E, et al. Evaluation of algorithms for photon depth of interaction estimation for the TRIMAGE PET component. EJNMMI Physics 2015;2:A13. https://doi.org/10.1186/2197-7364-2-S1-A13.
18. Green MV, Ostrow HG, Seidel J, Pomper MG. Experimental Evaluation of Depth-of-Interaction Correction in a Small-Animal Positron Emission Tomography Scanner. Mol Imaging 2010;9:311–8.
19. Lecoq P. Development of new scintillators for medical applications. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2016;809:130–9. https://doi.org/10.1016/j.nima.2015.08.041.